On sale

CS7600 Roller Cleaning Plate

Sale price

Regular price £38.40
( / )

The rollers on your CS7600 need cleaning every 1000 cycles. You will need this cleaning plate to carry out this maintenance.

To clean the scanner rollers, follow these steps:

Note: Cleaning the rollers requires using the size 4 insertion slot.

1: If the size 4 insertion slot is not installed, replace the insertion slot with the size 4 insertion slot.

2: Press the scanner’s left function button to access the Settings menu.

3: Select Rollers Cleaning.

4: Press "blue tick" the middle function button to confirm the Rollers Cleaning message. The Rollers Cleaning screen appears.

5: Peel the adhesive protective sheet from the rollers cleaning plate.

6: Insert gently the rollers cleaning plate into the size 4 insertion slot with the adhesive side first and the black stripe facing downwards until the rollers draw the rollers cleaning plate in.

7: Remove the roller cleaning plate from the insertion slot.

Repeat this process with the same cleaning plate 3 times to ensure the rollers are thoroughly cleaned.


 

 

Click here to be notified by email when CS7600 Roller Cleaning Plate becomes available.

This site is protected by hCaptcha and the hCaptcha Privacy Policy and Terms of Service apply.

This site is protected by hCaptcha and the hCaptcha Privacy Policy and Terms of Service apply.

The rollers on your CS7600 need cleaning every 1000 cycles. You will need this cleaning plate to carry out this maintenance.

To clean the scanner rollers, follow these steps:

Note: Cleaning the rollers requires using the size 4 insertion slot.

1: If the size 4 insertion slot is not installed, replace the insertion slot with the size 4 insertion slot.

2: Press the scanner’s left function button to access the Settings menu.

3: Select Rollers Cleaning.

4: Press "blue tick" the middle function button to confirm the Rollers Cleaning message. The Rollers Cleaning screen appears.

5: Peel the adhesive protective sheet from the rollers cleaning plate.

6: Insert gently the rollers cleaning plate into the size 4 insertion slot with the adhesive side first and the black stripe facing downwards until the rollers draw the rollers cleaning plate in.

7: Remove the roller cleaning plate from the insertion slot.

Repeat this process with the same cleaning plate 3 times to ensure the rollers are thoroughly cleaned.


 

 

What's New?

Revolutionising Dental Imaging with Acteon XMIND Prime2 CBCT

Revolutionising Dental Imaging with Acteon XMIND Prime2 CBCT

In today’s rapidly advancing dental industry, the integration of versatile imaging systems is paramount for executing successful dental procedures. 360...

New Carestream 3D Viewing Software: CS 3D Suite 3.10.38

New Carestream 3D Viewing Software: CS 3D Suite 3.10.38

Introducing the New 3D Viewing Software Upgrade: Boost Your Dental Practice Instruction Video: Here   Discover the latest 3D imaging...

The Carestream CS3800 has been rebranded the Dexis IS3800W

The Carestream CS3800 has been rebranded the Dexis IS3800W

Carestream Intra Oral Scanning business has been bought by Envista holdings corporation. The Carestream CS3800 has been rebranded the Dexis...

Revolutionising Dental Imaging with Acteon XMIND Prime2 CBCT

Revolutionising Dental Imaging with Acteon XMIND Prime2 CBCT

In today’s rapidly advancing dental industry, the integration of versatile imaging systems is paramount for executing successful dental procedures. 360...

New Carestream 3D Viewing Software: CS 3D Suite 3.10.38

New Carestream 3D Viewing Software: CS 3D Suite 3.10.38

Introducing the New 3D Viewing Software Upgrade: Boost Your Dental Practice Instruction Video: Here   Discover the latest 3D imaging...

The Carestream CS3800 has been rebranded the Dexis IS3800W

The Carestream CS3800 has been rebranded the Dexis IS3800W

Carestream Intra Oral Scanning business has been bought by Envista holdings corporation. The Carestream CS3800 has been rebranded the Dexis...